Reduction of CT artifacts caused by metallic implants.

Willi A. Kalender,R. Hebel,J. C. W. Ebersberger

Medicine
Reduction (mathematics)
Implant
1987
A technique to reduce metallic implant artifacts on computed tomography scans is presented. The implant boundaries are determined semiautomatically; the missing projection data are replaced by linear interpolation. The complete procedure requires 1-2 minutes per scan. Images with greatly improved quality were obtained in the presence of surgical clips and pelvic implants; success is limited in highly structured regions, such as the facial skull.